[Zurück]


Beiträge in Tagungsbänden:

D. Pollreisz, N. TaheriNejad:
"Reliable Respiratory Rate Extraction using PPG";
in: "2020 IEEE 11th Latin American Symposium on Circuits & Systems (LASCAS)", herausgegeben von: IEEE; IEEE, San Jose, Costa Rica, 2020, ISBN: 978-1-7281-3427-7, S. 1 - 4.



Kurzfassung englisch:
Wearable electronics enable a new look into the health of individuals in a fashion that was never possible before. However, many reliable methods for measuring Respiratory Rate (RR) require wearing gadgets that are impractical in a normal daily life setup. On the other hand, more practical methods, which are less intrusive, are often less reliable. Extracting RR using Photoplethysmogram (PPG) signals is one of the methods in the latter group. A major challenge for this method is the movement artifact, which leads to wrong estimation of RR or failure in its calculation. In this work, we propose a new algorithm, Smart Fusion of Frequency Domain Peak (SFFDP), that outperforms existing algorithm by at least 37% improvement in terms of reliability; i.e., average error, Standard Deviation (STD), and Figure of Merit (FoM). This method does not require any signal other than PPG. Therefore, it can be used in a wide range of wearable devices, such as smart watches, without any hardware additions.


"Offizielle" elektronische Version der Publikation (entsprechend ihrem Digital Object Identifier - DOI)
http://dx.doi.org/10.1109/LASCAS45839.2020.9068993

Elektronische Version der Publikation:
https://publik.tuwien.ac.at/files/publik_293730.pdf


Erstellt aus der Publikationsdatenbank der Technischen Universität Wien.