Zeitschriftenartikel:
B. Povazay, A. Unterhuber, B. Hermann, H. Sattmann, H. Arthaber, W. Drexler:
"Full-field time-encoded frequency-domain optical coherence tomography";
Optics Express,
vol. 14
(2006),
S. 7661
- 7669.
Kurzfassung englisch:
Ultrahigh axial resolution surface profiling as well as volumetric optical imaging based on time encoded optical coherence tomography in the frequency domain without any mechanical scanning element is presented. A frequency tuned broad bandwidth titanium sapphire laser is interfaced to an optical microscope (Axioskop 2 MAT, Car Zeiss Meditec) that is enhanced with an interferometric imaging head. The system is equipped with a 640 x 480 pixel CMOS camera, optimized for the 800 nm wavelength tuning range for transmission and reflection measurements of a microscopic sample. Sample volume information over 1.3 x 1 x 0.2 mm3 with ~3 µm axial and ~4 µm transverse resolution in tissue is acquired by a single wavelength scan over more than 100 nm optical bandwidth from <760 to >860 nm with 128-2048 equidistant optical frequency steps with an acquisition time of 1 to 50ms per step. Topography and tomography with a signal to noise ratio of 83 dB is demonstrated on test surfaces and biological specimen respectively. This novel OCT technique promises to enable high speed, three dimensional imaging by employing high frame rate cameras and state of the art tunable lasers in a mechanically stable environment, due to lack of moving components while reducing the intensity on the sample.
Online-Bibliotheks-Katalog der TU Wien:
http://aleph.ub.tuwien.ac.at/F?base=tuw01&func=find-c&ccl_term=AC06586237